Six steps from jiggle to radius
Brownian motion
Thermal kicks from solvent molecules make each particle random-walk.
Interference
Coherent light scattered by all particles adds up into a speckle pattern.
Intensity trace
A photon counter watching one speckle sees the intensity fluctuate.
Autocorrelation
How similar is the signal to itself after a delay τ?
Decay rate
The correlation decays exponentially with rate Γ.
Stokes–Einstein
Diffusion coefficient becomes an equivalent-sphere radius.
The scattering vector sets the length scale
Light with wavevector \(\mathbf k_i\) enters the sample, and the detector collects light with wavevector \(\mathbf k_s\) at angle \(\theta\). For quasi-elastic scattering \(|\mathbf k_i| = |\mathbf k_s| = 2\pi n/\lambda_0\), and the momentum transfer is
Here \(n\) is the refractive index of the solvent and \(\lambda_0\) is the vacuum wavelength. DLS probes motion over distances of order \(2\pi/q\), a few hundred nanometres. A particle has to move about that far before the speckle it contributes to has changed.
Commercial instruments often detect at 173° (non-invasive backscatter) to shorten the optical path and reduce multiple scattering, or at 90° (classic goniometer), or at small forward angles to see large particles. Changing \(\theta\) changes \(q\), so it changes the decay rate, but for pure translational diffusion it must not change the size you calculate.
Brownian motion and Stokes–Einstein
A free particle in a viscous fluid performs a random walk. For times much longer than the momentum relaxation time \(m/(6\pi\eta R)\), which is nanoseconds or less for colloids, its mean-squared displacement grows linearly in time:
The Einstein relation \(D = k_BT/f\) links diffusion to the friction coefficient \(f\). For a sphere with stick boundary conditions, Stokes gives \(f = 6\pi\eta R\), hence
\(R_h\) is the hydrodynamic radius: the radius of a hard sphere that diffuses at the same speed. It includes the solvation shell, adsorbed polymer, ligands and counter-ion cloud, so it is usually larger than a TEM or crystallographic radius. For non-spherical objects it is an effective value only.
Viscosity is the parameter people most often get wrong. It depends strongly on temperature: water loses about 2 % of its viscosity per kelvin near room temperature. A convenient empirical form for water is
What moves how far?
Root-mean-square 1D displacement \(\sqrt{2D\tau}\) in water at 25 °C, over the time it takes the speckle to decorrelate.
| Diameter | D | 1/Γ at 173° | xrms in 1 ms |
|---|
The diffusion time scales linearly with size, so a thousandfold range in diameter maps onto three decades in τ. That is why correlators use logarithmically spaced lag times.
From a flickering signal to a correlation function
The detector records photon counts, which give the intensity \(I(t)\). The correlator computes the normalised intensity autocorrelation function
What we want physically is the field autocorrelation function, which is the dynamic structure factor of the sample:
If the scattered field is a sum of many independent contributions, the central limit theorem makes it a complex Gaussian random variable. Then the fourth moment in \(g_2\) factorises and we get the Siegert relation:
The baseline \(B\) is 1 for an ideal measurement. The coherence factor \(0 < \beta \le 1\) (the intercept) depends on how many speckles the detector sees. Single-mode fibre optics give \(\beta\) near 0.9; large pinholes average several speckles and reduce it.
Why an exponential? For dilute, non-interacting particles, only self-terms survive the average:
The second step uses the fact that Brownian displacements are Gaussian. The decay rate is therefore
Note the factor of 2: the intensity correlation decays twice as fast as the field correlation. Mixing up \(\Gamma\) and \(2\Gamma\) gives sizes off by exactly a factor of two, which is a classic self-written-fit error.
Homodyne vs heterodyne. Everything above assumes only scattered light reaches the detector (homodyne). If a strong static reference field is mixed in, from a local oscillator, from wall flare or from immobile scatterers in a gel, the signal becomes linear in \(g_1\):
Then the apparent decay rate is \(\Gamma\) instead of \(2\Gamma\). Unintended partial heterodyning from a scratched cuvette is a real source of error.
Correlation function explorer
Change the sample and instrument and watch \(g_2(\tau)-1\) respond. The size distribution is lognormal and intensity-weighted. The dashed curve is the second-order cumulant approximation that instruments report as Z-average and PDI.
Things to try: raise the temperature and see the decay speed up through both T and η(T). Switch dust on and see the long-time shoulder that no single exponential explains. Widen σ and watch PDI rise roughly as σ².
Cumulants, Z-average and the inverse problem
A real sample contains a distribution of decay rates \(G(\Gamma)\), weighted by how much light each species scatters:
This is a Laplace transform. Inverting it from noisy data is ill-posed: very different \(G(\Gamma)\) produce correlation functions that agree within noise. Everything in DLS analysis is a strategy for coping with that.
Method of cumulants (Koppel, 1972)
Expand the logarithm of \(g_1\) in powers of \(\tau\):
Frisken (2001) recommends fitting the intensity data directly, keeping baseline and intercept free, instead of fitting a log-linearised form:
The mean rate converts into the Z-average diameter. In the Rayleigh regime it is the intensity-weighted harmonic mean:
Reading PDI
| PDI | Typical meaning |
|---|---|
| < 0.05 | Near-monodisperse, e.g. latex size standards |
| 0.05–0.1 | Narrow; cumulant result is trustworthy |
| 0.1–0.4 | Moderately polydisperse; look at a distribution too |
| > 0.4–0.5 | Broad or multimodal; Z-average alone is not meaningful |
For a narrow lognormal distribution of width \(\sigma\), PDI is approximately \(\sigma^2\). The relative standard deviation of the decay-rate distribution is \(\sqrt{\mathrm{PDI}}\).
Distribution methods
To get \(G(\Gamma)\) itself, discretise eq. (10) on a grid of \(\Gamma_j\) and solve a constrained, regularised least-squares problem:
CONTIN (Provencher, 1982) uses a smoothness penalty \(L\) on the second derivative and chooses \(\alpha\) statistically. NNLS drops the penalty and keeps only non-negativity, which gives spikier results. As a rule of thumb, two populations need to differ in size by a factor of about 3 or more before they separate reliably, even with excellent data.
Always report the method, the regularisation settings and whether the plotted distribution is weighted by intensity, volume or number.
Intensity, volume and number are three different answers
DLS measures an intensity-weighted distribution. For particles much smaller than the wavelength (Rayleigh regime, roughly \(d < \lambda/10\)), scattered intensity per particle scales with the square of its polarisable volume:
So a small number of large particles can dominate the signal. Converting to volume or number distributions needs the particle refractive index and Mie theory for larger particles (the form factor \(P(qR)\) modifies the \(d^6\) law) and it amplifies noise. Treat converted number distributions as rough guides.
The widget uses the Rayleigh law with two populations. Set the number ratio so that both populations scatter equally, and notice how lopsided the number picture becomes.
Pitfalls, ranked by how badly they hurt
Dust and aggregates
With \(d^6\) weighting, a few large particles swamp the signal. Filter solvents (0.02–0.2 µm), filter or centrifuge samples when the particles allow it, rinse cuvettes with filtered solvent, and watch for spikes in the count-rate trace.
Multiple scattering
In turbid samples light scatters more than once, decorrelates faster, and particles look smaller. Signs are a low intercept and a size that drifts with concentration. Dilute, use backscatter optics or move to cross-correlation (3D-DLS) or DWS.
Wrong η, n or T
Size is proportional to \(T/\eta\) and to \(q^2\). Use the viscosity of the actual dispersant (buffers with glycerol, sucrose or high salt differ from water), the solvent refractive index, and let the cell equilibrate thermally before the first run.
Interactions and concentration
Stokes–Einstein holds for infinite dilution. At finite concentration you measure a collective diffusion coefficient. Run a dilution series and extrapolate to \(c \to 0\).
Low intercept, noisy baseline
An intercept well below the instrument's usual value points to weak scattering, multiple scattering or fluorescence. A baseline that does not settle at 1 points to drift, sedimentation or number fluctuations of a few large objects.
Sedimentation and large sizes
Above roughly a micrometre, particles settle through the beam, only a few are in the scattering volume at once, and the upper limit of the method is near. Check with microscopy or laser diffraction.
Over-reading distributions
A second peak from one run is not a result. It must be reproducible across runs, stable across settings, and consistent with the correlation function residuals.
Absorbing or fluorescent samples
Absorption heats the sample locally and fluorescence adds uncorrelated light that lowers β. Change wavelength or add a filter.
Non-spheres and soft objects
Rods, discs and swollen gels have no unique radius. Report \(R_h\) as an equivalent-sphere value and consider angle-dependent or depolarised measurements.
A measurement protocol worth copying into your lab book
- Validate the instrument with a certified latex or silica standard in the relevant size range before a measurement series.
- Prepare clean dispersant by filtering it and measuring it alone. Count rate should be low and stable.
- Choose concentration by dilution series: size and PDI must be independent of concentration within error.
- Equilibrate temperature for several minutes after inserting the cell and record the actual sample temperature.
- Acquire several runs (at least 3 to 5, each long compared with the slowest decay) and inspect the count-rate trace of every run.
- Inspect raw g₂ for intercept, baseline, smoothness and fit residuals before looking at any size.
- Report fully: T, dispersant η and n, λ₀, θ, concentration, analysis method, weighting, mean ± SD over replicates. ISO 22412:2017 lists what to include.
Diagnosing a correlation function by eye
| What you see | Likely cause |
|---|---|
| Single smooth sigmoid on log τ | Good data; monomodal sample |
| Two separate shoulders | Bimodal sample, e.g. monomer plus aggregate |
| Long tail, baseline never reached | Dust, large aggregates or sedimentation |
| Baseline below 1 (negative g₂−1) | Drift in count rate during the run |
| Intercept much lower than usual | Weak scattering, multiple scattering, fluorescence |
| Noise at short τ | Afterpulsing of the detector or too few counts |
| Decay that does not scale as q² | Non-diffusive motion: rotation, internal modes, flow, convection |
Beyond the single-angle Z-average
Multi-angle DLS
For translational diffusion, \(\Gamma/q^2\) must be constant. Plot \(\Gamma\) against \(q^2\): a straight line through the origin confirms diffusion, and deviations reveal internal dynamics or polydispersity combined with form-factor effects.
Concentration dependence
Interactions renormalise diffusion. The interaction parameter \(k_D\) contains the second virial coefficient and hydrodynamic terms and is widely used to screen protein formulations.
Shape from Rg/Rh
Combine with static light scattering. The ratio \(\rho = R_g/R_h\) is about 0.775 for a homogeneous hard sphere, about 1.5 for a random coil and above 2 for elongated rods. Hollow shells give values near 1.
Depolarised DLS
Optically anisotropic particles depolarise light. The VH signal contains rotation:
Fitting several angles separates translational and rotational diffusion.
Non-ergodic samples
In gels and glasses the particles cannot explore all configurations. A time average at one spot is not the ensemble average. Use sample rotation, multi-speckle cameras or the Pusey–van Megen treatment.
Related methods
Diffusing-wave spectroscopy (turbid samples), X-ray photon correlation spectroscopy (nanometre length scales), differential dynamic microscopy (microscope camera), and electrophoretic light scattering (zeta potential) share the same correlation ideas.
100 nm polystyrene latex in water
Conditions: \(\lambda_0 = 632.8\) nm, \(\theta = 173°\), water at 25 °C (\(n = 1.330\)). Every number below is computed on this page from the equations above.
Reading it back
–
Sensitivity check
Because \(d \propto T q^2/(\eta\,\Gamma)\), relative errors add up:
–
Quick formulas
| Boltzmann constant | kB = 1.380649 × 10⁻²³ J/K |
| kBT at 25 °C | 4.116 × 10⁻²¹ J |
| Water η at 20 / 25 / 37 °C | – |
References
- B. J. Berne and R. Pecora, Dynamic Light Scattering: With Applications to Chemistry, Biology, and Physics, Wiley (1976); Dover reprint (2000). The standard monograph.
- D. E. Koppel, “Analysis of macromolecular polydispersity in intensity correlation spectroscopy: the method of cumulants,” J. Chem. Phys. 57, 4814 (1972).
- S. W. Provencher, “CONTIN: a general purpose constrained regularization program for inverting noisy linear algebraic and integral equations,” Comput. Phys. Commun. 27, 229 (1982).
- B. J. Frisken, “Revisiting the method of cumulants for the analysis of dynamic light-scattering data,” Appl. Opt. 40, 4087 (2001).
- P. N. Pusey and W. van Megen, “Dynamic light scattering by non-ergodic media,” Physica A 157, 705 (1989).
- J. Stetefeld, S. A. McKenna and T. R. Patel, “Dynamic light scattering: a practical guide and applications in biomedical sciences,” Biophys. Rev. 8, 409 (2016).
- W. Schärtl, Light Scattering from Polymer Solutions and Nanoparticle Dispersions, Springer (2007).
- ISO 22412:2017, Particle size analysis: Dynamic light scattering (DLS).